Sterling Acuatic Vacuum 4 Steriliser:
Fast cycle time • Simple to use • Low maintenance and running costs • Full colour touchscreen • Inbuilt "Securilog" data logger • 5 year parts guarantee • Manufactured in the UK • Excellent value for money • Outstanding reliability • Supplied with a free Helix tester • Advanced vacuum sterilisation
Specification:
Dimensions:
-
Width: 345 mm
-
Height: 270 mm
-
Length: 440 mm
-
Weight: 16 Kg
-
Chamber Diameter: 150 mm
-
Chamber Length: 230 mm
-
Chamber Volume: 4 litres
Trays:
-
Number: 3
-
Width: 125 mm
-
Length: 225 mm
-
Max load per tray: 1 Kg
Clearances:
-
Top and sides: 50mm
-
Rear: None required
Cycle Type:
Cycle Time:
Sterilising Parameters:
-
Temperature: 134°C to 138°C
-
Time: 3 minutes minimum
Limits:
-
Design pressure: -1 to 3 bar
-
Design temperature: 2°C to 144°C
Display resolution:
-
Temperature: 0.1°C
-
Pressure: 0.01 bar
Display accuracy:
-
Between 75°C and 150°C: ±2°C
-
Between 134°C and 138°C: +0 / -1.5°C
Electrical:
-
Voltage: 230 V ac ± 10%
-
Current: Variable up to 13 A
-
Power: Variable up to 3 kW
-
Frequency: 50 Hz
-
Plug top fuse: 13 A
-
Internal fuse: 1 A
-
Energy used per cycle: 0.1 to 0.2 kWh*
Water:
-
Water used per cycle: 100 to 200 ml*
-
Clean water reservoir volume: 1 litre
-
Used water reservoir volume: 1 litre
Filling:
-
Manual fill through front filling spout
Draining:
-
Manual drain with plug in hose to front drain connection
Cycle recording:
-
Type: Inbuilt 'Securilog' data logger
-
Recording media: SD memory card
-
Capacity: Lifetime of steriliser
* Cycle time and electrical and water usage is dependent on steriliser model, loading and cycle start temperature. Cycle times quoted above do not include drying which can be set to between 0 and 60 minutes.